Model Number N/A
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and found power supply board defective, replaced the power supply board and fault corrected.Then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
 
Event Description
It was reported that during routine check, the power supply board is defective in the cs100 intra-aortic balloon pump (iabp).There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during routine check, the power supply board is defective in the cs100 intra-aortic balloon pump (iabp).There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11515313
MDR Text Key240665944
Report Number2249723-2021-00535
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public10607567107745
Combination Product (y/n)N
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/20/2021
Initial Date FDA Received03/18/2021
Supplement Dates Manufacturer Received05/24/2021
Supplement Dates FDA Received05/24/2021
Patient Sequence Number1
-
-