DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit and found power supply board defective, replaced the power supply board and fault corrected.Then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
|
|
Event Description
|
It was reported that during routine check, the power supply board is defective in the cs100 intra-aortic balloon pump (iabp).There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during routine check, the power supply board is defective in the cs100 intra-aortic balloon pump (iabp).There was no patient involvement, and no adverse event reported.
|
|