The patient underwent a thrombectomy procedure in the right middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68) and a neuron max 6f 088 long sheath (neuron max).Afterwards, the physician noticed some light spots via a computed tomography scan (ct scan), possibly due to some scraps peeled off from the distal markerband.There was no report of an adverse effect to the patient.
|
Evaluation of the returned distal segment of the ace68 revealed that the marker band was undamaged and intact on the distal tip of the catheter.Therefore, the reported scraps peeled off from the distal marker band could not be confirmed.The ace68 had been cut and only the distal segment was returned for evaluation.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|