• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR WEST; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR WEST; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX25RW
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during out of box, there was no connection port in the wrapping.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307) type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #3: 3331 - analysis of production records investigation findings: 3259 - improper physical structure investigation conclusions: 4307 - cause traced to component failure the affected sample was inspected upon receipt to confirm a broken blood inlet port on the oxygenator.A representative retention sample was reviewed and found to have no damage oxygenator, specifically in the area of the blood inlet.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 19, 2021.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX25 OXYGENATOR WEST
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key11525758
MDR Text Key245524303
Report Number1124841-2021-00057
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450110
UDI-Public(01)00699753450110
Combination Product (y/n)N
PMA/PMN Number
K153213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number3CX*RX25RW
Device Catalogue NumberN/A
Device Lot NumberYD16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/23/2021
Initial Date FDA Received03/19/2021
Supplement Dates Manufacturer Received03/25/2021
04/26/2021
Supplement Dates FDA Received04/15/2021
05/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-