• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-3-A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Arrest (1762); Respiratory Arrest (4461)
Event Date 02/24/2021
Event Type  Death  
Manufacturer Narrative
The device was received for evaluation and successfully passed the prime and alarms test confirming proper functionality of the device and sensors.Available log files were retrieved and analyzed which showed device performance was without incident and was unremarkable.A review of the device history record confirmed the product was released meeting all product design specifications and quality criteria.All information supports the device was functioning as designed and intended with no malfunction.The nxstage system one user guide supplement for solo home hemodialysis outlines risks, warnings, and requirements for the solo home hemodialysis patient.(b)(4).
 
Event Description
A report was received on 25 feb 2021 from the home therapy nurse (htn) of an (b)(6) male patient with a medical history of multiple comorbidities and end stage renal disease, approved for performing solo home hemodialysis therapy, stating the patient was found unresponsive on (b)(6) 2021.Additional information was received on 26 feb 2021 from the htn who stated that emergency services were contacted and the patient was pronounced on scene.Per the htn, the available treatment information was unremarkable and the medical doctor suspects the cause of death to be cardiopulmonary arrest unrelated to nxstage product or therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
nxstage medical, inc
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key11529464
MDR Text Key241060034
Report Number3003464075-2021-00011
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122051
Number of Events Reported123
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-3-A
Device Catalogue NumberCHRONIC HI-FLOW CYCLER, ROHS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/25/2021
Initial Date FDA Received03/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age86 YR
Patient Weight73
-
-