|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/19/2021 |
Event Type
malfunction
|
Event Description
|
The customer reported that the unit's touchscreen was out of alignment.The customer reported there was no patient involvement.The manufacturer's product support engineer (pse) evaluated the unit and confirmed the reported issue.During the oxygen concentration accuracy test it was found that the values were off.Pse replaced the unit's touchscreen to resolve the reported issue.To resolve the oxygen concentration testing issue, the flow sensor assembly was replaced.The unit passed required performance verification tests per philips standards.
|
|
Manufacturer Narrative
|
The reported issue was during preventive maintenance, which is outside of clinical use and, therefore, not reportable.The touchscreen assembly was returned for failure investigation.Visual inspection of the touchscreen assembly revealed no evidence of damage or contamination.The ul_lr and ur_ll resistance were out of specifications.
|
|
Search Alerts/Recalls
|
|
|