|
Model Number REPLY DR |
Device Problem
Pacing Intermittently (1443)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/19/2021 |
Event Type
malfunction
|
Event Description
|
During a replacement procedure due to normal battery depletion of the subject pacemaker, a temporary lead for a patient without intrinsic beat was inserted before the intervention.At that time, ecg waveforms that seem to be a ventricular pacing failure were observed on the monitor.The pacemaker was checked, but the event was not reproduced.Since the measurement data of the pacemaker was the same as before, the pacemaker replacement intervention was completed.However, since the cause could not be identified, it was still decided to follow up in the ward with the temporary lead inserted even after the replacement.The physician requested an analysis of the subject pacemaker.
|
|
Event Description
|
During a replacement procedure due to normal battery depletion of the subject pacemaker, a temporary lead for a patient without intrinsic beat was inserted before the intervention.At that time, ecg waveforms that seem to be a ventricular pacing failure were observed on the monitor.The pacemaker was checked, but the event was not reproduced.Since the measurement data of the pacemaker was the same as before, the pacemaker replacement intervention was completed.However, since the cause could not be identified, it was still decided to follow up in the ward with the temporary lead inserted even after the replacement.The physician requested an analysis of the subject pacemaker.
|
|
Search Alerts/Recalls
|
|
|