• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US UNK - IMPLANT; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US UNK - IMPLANT; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number UNK - IMPLANT
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Joint Laxity (4526); Unspecified Tissue Injury (4559)
Event Type  Injury  
Manufacturer Narrative
This report is for an unknown implant.Part and lot number are unknown.Without the specific part number; the udi number, 510-k number and manufacturing site name are unknown.Complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/ or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
This file is a review of the following journal article: russo, r., et al (2014) risk factors for recurrent shoulder dislocation arthroscopically managed with absorbable knotless anchors.Advances in orthopedic surgery, pages 1-7, (italy).The study emphasizes on the evaluation of the clinical outcome and risk factors for recurrent dislocation after arthroscopic stabilization with absorbable knotless anchor.The patients evaluated on course of this study: 197 patients.The article describes the following procedure: shoulder stabilization.The device involved was: unknown bioknotless anchor.Complications described: 10 cases of dislocation recurrence.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned; the availability of the device is unknown, therefore unavailable for a physical evaluation.This complaint file was opened to document complaints derived through a journal article review.The journal article review indicated depuy mitek product failure(s).Multiple attempts were done to obtain more information from the author; however, no response was received.It is unknown if complaints derived from this journal article were previously reported and documented in the depuy mitek complaint system at the time of occurrence as no product code/lot number information was provided to perform the search.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported problem.Since no lot number was provided, a manufacturing record evaluation or sterile load review could not be conducted.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - IMPLANT
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key11555760
MDR Text Key242012312
Report Number1221934-2021-00939
Device Sequence Number1
Product Code MAI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK - IMPLANT
Initial Date Manufacturer Received 03/08/2021
Initial Date FDA Received03/23/2021
Supplement Dates Manufacturer Received04/05/2021
Supplement Dates FDA Received04/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-