• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Model Number 72203967S
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case: (b)(4).
 
Event Description
It was reported that during surgery, the camera head had no image.A backup device was available to complete the procedure with no delay or patient injuries.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was returned for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed the device was not recognized when plugged in.The complaint was confirmed and a root cause has been associated with an electrical component failure.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include a damaged connector.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11560761
MDR Text Key243144211
Report Number1643264-2021-01127
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2021
Initial Date FDA Received03/24/2021
Supplement Dates Manufacturer Received04/09/2021
Supplement Dates FDA Received04/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-