The complaint wire was not returned, thus visual inspection could not be undertaken.No futher clinical information or imaging was available, thus the investigation was limited to internal reviews.Review of the device history records confirmed that the product was supplied in conformance with ortho solutions specifications.A wire from the same batch was previously tested under worst-case conditions, and the device performed as intended and fulfilled the acceptance criteria with no deviations.The root cause for the complaint condition could not be ascertained following the investigation, however the manufacture of the wire was not likely to be the contributory cause for the issue.The complaint does not suggest any new or emerging risks with the device when utilised in accordance to its intended use.
|
During an orif ankle procedure, where the patient required two cannulated screws bi-malleolar, it was reported that the k-wire broke upon insertion and retained in the patient.The remnant portion of the wire was discarded by the healthcare facility.The surgical procedure proceeded successfully, and no patient injury was reported.No further complications which required medical intervention were reported.
|