Catalog Number UNK HIP ACET CUP ASR
Device Problems Noise, Audible (3273); Illegible Information (4050)
Patient Problems Foreign Body Reaction (1868); Pain (1994)
Event Date 01/26/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).No 510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The asr/corail hip was revised due to pain.The surgeon revised the hip, and when he got in he found that then taper on stem has been badly worn out by the asr xl and there was huge metal wear.He femoral component was damaged as seen, but well fixed.Cup also well fixed, severe metallosis in the joint.Both components changed.Postoperative uneventful course.The lot numbers are worn out, and can¿t be read on the implants.Doi: (b)(6) 2009, dor: (b)(6) 2021, left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : this hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : previous investigations that have included device history reviews since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing, or inspection.Therefore, no device history record (dhr) review for this individual asr component will be carried out at this point in time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP ACETABULAR CUP ASR
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key11567908
MDR Text Key242190773
Report Number1818910-2021-06055
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACET CUP ASR
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/22/2021
Initial Date FDA Received03/25/2021
Supplement Dates Manufacturer Received05/24/2021
Supplement Dates FDA Received05/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNK HIP FEMORAL STEM CORAIL; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNK HIP FEMORAL STEM CORAIL
Patient Outcome(s) Required Intervention;
-
-