One needle holder and one guidewire with holder was returned for evaluation.Packaging and all other components were not returned.Customer report of issue with guidewire was confirmed during evaluation.Damage was found on the j tip area of the returned guidewire.The guidewire coil was found to be damaged and uneven at two locations on the j tip area.No other visible damage to the guidewire was observed.Additionally, the customer report of issue with needle holder was not confirmed.No visible damage was observed from the returned needle holder during visual examination.It was able to interlock the needle holder, and hold a lab needle without any problem.Visual examinations were performed under microscope at 10x magnification.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.In this instance, damage was found on the j tip area of the returned guidewire.The guidewire coil was found to be damaged and uneven at two locations on the j tip area.In this event, there were no patient complications noted as the issue was found before use.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|