GUANGDONG HAIOU MEDICAL APPARATUS CO., LTD AIOU NEEDLE RETRACTABLE SAFETY SYRINGE WITH NEEDLE 23GX1"; SYRINGE, ANTISTICK Back to Search Results
Lot Number 20JC2
Device Problems Break (1069); Leak/Splash (1354); Material Separation (1562)
Patient Problem Insufficient Information (4580)
Event Date 03/20/2021
Event Type  malfunction  
Event Description
Syringe broke apart while covid-19 vaccine was being administered.Needle and hub separated during vaccine administration and vaccinator and patient were sprayed with vaccine.Needle and hub remained in patient's arm despite no longer being attached to syringe.Vaccinator carefully removed needle and hub from patient's arm and disposed of it in sharps container.Syringe type: aiou needle retractable safety syringe with needle 23gx1" lot 20jc2.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIOU NEEDLE RETRACTABLE SAFETY SYRINGE WITH NEEDLE 23GX1"
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
GUANGDONG HAIOU MEDICAL APPARATUS CO., LTD
MDR Report Key11574212
MDR Text Key242752507
Report NumberMW5100266
Device Sequence Number1
Product Code MEG
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Unknown
Type of Report Initial
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Lot Number20JC2
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/25/2021
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N