There is a photo toxicity message on the screen, and photo toxicity is explained in the device user manual.The incident was not reported by customer for 26 days after the incident.The stellaris system issued a warning message las07 - ensure that the eye safety filter is installed (endo mode) and operating room personnel have proper eye protection - during the timeframe of the incident.The root cause is inconclusive.The device history was reviewed and found to meet manufacturing specification.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
|