Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.The investigation was performed considering complaint description, cs reports and system history.No logs were provided; therefore, a deeper evaluation of this event could not be performed due to the lack of information.According to the customer, the system freezes during operation and no further use of the system could be done.The customer received the message "the application has a serious memory management or internal communication problem" that was displayed at the monitor.The customer did a "forced shut down" of the system by using the reset button.After pressing the on button, the pc did not start, and the system could not be used.When the customer service engineer (cse) arrived onsite and switched the system on, the system started but the error message "internal imaging system error occurred" appears at the monitor.Some software activities were not available.The cse reloaded the software, and the system could be completely recovered afterwards.The patient database was deleted in agreement with the customer.No parts were exchanged and no systematic issue was identified.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
|