The patient was undergoing a thrombectomy procedure in the internal iliac artery (iia) using a penumbra system aspiration pump max 220 (pump max) and an indigo system aspiration catheter 6 (cat6).During the procedure, the physician completed a pass using the cat6 and pump max.Next, the physician removed the cat6 and the pump max was turned off.Afterwards, the physician advanced the cat6 for another pass and turned on the pump max.However, the pump max would not turn on.Subsequently, the pump max turned on after several attempts were made.The procedure was completed using the same pump max and the same cat6.There was no report of an adverse effect to the patient.
|
Please note that the following sections were incorrectly reported on the initial mfr report and are being corrected on this follow-up #01 mfr report:3005168196-2021-00630 1.Section h.Box 5.Labeled for single use? 2.Section h.Box 8.Usage of device.Evaluation of the returned pump max revealed a functional device.During the functional testing, the pump max was able to power on and produce a vacuum pressure within specification.The pump was turned on and turned off several times and no issue found.The reported complaint could not be determined.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|