Model Number 0605510
Device Problems Component Missing (2306); No Apparent Adverse Event (3189); Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event Description
It was reported that during a port placement, the pressure indication was allegedly missing on the patient card.There was no reported patient injury.
 
Manufacturer Narrative
H10: the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the investigation is inconclusive for the reported pressure indication missing on patient cards, as the device was not returned for evaluation.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement, the pressure indication was allegedly missing on the patient card.There was no reported patient injury.
 
Event Description
It was reported that during a port placement, the pressure indication was allegedly missing on the patient card.There was no reported patient injury.
 
Manufacturer Narrative
Additional information was received and the reportability was reassessed as malfunction.Manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is confirmed for the reported customer feedback issue as a report has been received but the description provided does not appear to relate to an adverse event.However, the investigation is unconfirmed for the reported pressure indication missing on patient card issue as for the non-power ports there were no pressure indication on the patient cards.Furthermore, a 25 psi maximum pressure was mentioned in the instruction for use.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Instruction for use and it states that, precautions: do not use a syringe smaller than 10 ml! infusion pressure greater than 25 psi (172 kpa) may damage blood vessels and viscus and is not recommended.Do not use a syringe smaller than 10 ml.Flushing occluded catheters with small syringes can create excessive pressure within the port system.The current baw0734901 rev.0 states that, warning: do not use a syringe smaller than 10 ml.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM LOW-PROFILE IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11593780
MDR Text Key243130712
Report Number3006260740-2021-01054
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741025969
UDI-Public(01)00801741025969
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K050310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0605510
Device Catalogue Number0605510
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2021
Initial Date FDA Received03/31/2021
Supplement Dates Manufacturer Received07/21/2021
01/08/2022
Supplement Dates FDA Received07/27/2021
01/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1