MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Model Number 72203967S
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the camera head was blacked out.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was returned to the designated complaint station for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed a dark screen display with no signal.The complaint has been confirmed and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event includes a damaged connector.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11598741
MDR Text Key243330252
Report Number1643264-2021-01237
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (y/n)N
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Initial Date Manufacturer Received 03/10/2021
Initial Date FDA Received03/31/2021
Supplement Dates Manufacturer Received04/27/2021
Supplement Dates FDA Received04/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-