• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP XTR CLIP DELIVERY SYSTEM; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP XTR CLIP DELIVERY SYSTEM; VALVE REPAIR Back to Search Results
Catalog Number CDS0602-XTR
Device Problems Component Misassembled (4004); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report clip cover split.It was reported that prior to use, the plastic protective clip cover was found not assembled over the clip.Additionally, the plastic protective cover was cut/slit.The clip delivery system (cds) was not used in the patient.The procedure was successfully completed with a new cds.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Event Description
N/a.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All information was investigated, and the reported clip cover protective pouch being torn resulting in the clip cover protective pouch not assembled over the clip appears to be related to a potential product issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP XTR CLIP DELIVERY SYSTEM
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11605967
MDR Text Key245528740
Report Number2024168-2021-02649
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2021
Device Catalogue NumberCDS0602-XTR
Device Lot Number00908U202
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/08/2021
Initial Date FDA Received04/01/2021
Supplement Dates Manufacturer Received05/04/2021
Supplement Dates FDA Received05/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-