Catalog Number 2400B |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the base plate of an exacta mix automated compounding device was contaminated; further described as microbial growth.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.A visual inspection was performed which noted the device had a physical damage to the display arm bracket; arm was not intact.The reported condition was verified as physical damaged to the base plate.The cause of the physical damage to the base plate was due to ingredient spills (contamination) that may have not been properly and immediately cleaned.The base plates are not serviceable and will be scrapped.Should additional relevant information become available, a supplemental report will be submitted.
|
|