Catalog Number 2400M
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the main module of an exacta mix automated compounding device was contaminated; further described as microbial growth.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed which noted the device had a stained bag pan.There was considerable dried, spilled ingredient or cleaning solution oxidation (green) stains on the bottom of main housing copper coating.Functional and electrical testing was performed with no issues noted.The reported issue of microbial growth was not verified; however, a stained bag pan/oxidation of the housing was verified.The cause of the staining/oxidation could not be determined; however, is most likely due to ingredient spills or cleaning solutions, normal wear and tear of unit being in use, or customer not keeping unit clean and dry.The staining/oxidation is not likely to cause harm as this occurred outside of the fluid path.The main module housing cover will be replaced to correct the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, MAIN MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11608246
MDR Text Key246114951
Report Number1416980-2021-01829
Device Sequence Number1
Product Code NEP
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2021
Initial Date Manufacturer Received 03/10/2021
Initial Date FDA Received04/02/2021
Supplement Dates Manufacturer Received04/12/2021
Supplement Dates FDA Received05/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-