SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.The complaint of fan malfunction and overheating was confirmed.The cooling fan for the ballast/power supply shuts down when unit is powered up causing overheating and errors.Fan malfunction is caused by a defective mcu pcb.The complaint investigation has concluded the cause of the failure to be a defective electronic component.  a review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that the light source 500 xl xenon fan did not start, so the device was overheating, that is all the information provided.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.The complaint of fan malfunction and overheating was confirmed.The cooling fan for the ballast/power supply shuts down when unit is powered up causing overheating and errors.Fan malfunction is caused by a defective mcu pcb.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11611265
MDR Text Key243831547
Report Number1643264-2021-01245
Device Sequence Number1
Product Code FFS
UDI-Device Identifier03596010643070
UDI-Public03596010643070
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200568
Device Catalogue Number72200568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2021
Initial Date Manufacturer Received 03/11/2021
Initial Date FDA Received04/02/2021
Supplement Dates Manufacturer Received04/01/2021
Supplement Dates FDA Received04/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-