• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5090S
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problems Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation as it remains in-situ and the lot number was not provided.The root cause is unable to be determined at this time.If any relevant additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that the end cap migrated.According to the report, the intended lengthening has been achieved to date and there was no report of patient impact or revision surgery.No additional information is available at this time.
 
Manufacturer Narrative
Additional information: h.6, h.7, h.9, h.10.Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Event Description
No new information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key11615119
MDR Text Key244918188
Report Number3006179046-2021-00220
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (y/n)N
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-5090S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/10/2021
Initial Date FDA Received04/05/2021
Supplement Dates Manufacturer Received03/10/2021
Supplement Dates FDA Received06/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1898-2020
Patient Sequence Number1
Patient Age7 YR
Patient Weight19
-
-