The patient was undergoing a thrombectomy procedure using a penumbra engine (engine).During the procedure, the engine was noticed to be cracked and not functioning properly.The crack was noticed during the start of the procedure when the four indicator lights would not initiate, and the vacuum pressure was sub-par.Therefore, the engine was not used for the remainder of the procedure.The procedure was completed using another pump.There was no report of an adverse effect to the patient.
|
Evaluation of the returned engine confirmed that vacuum could not be achieved.Evaluation also revealed that the black lid to the engine was separated from the housing.This is likely the reported crack.During the functional test, the engine was powered on and vacuum could not be achieved within specification, no vacuum indicator lights were illuminated, and engine did not vibrate on the table.The engine housing was opened, and the vacuum tubing barb was fractured off the sensor.This likely contributed to the engine not achieving vacuum within specification.Further evaluation revealed that the one of the screws to the lid was loose an another one of the screws was fractured from the lid.This damage likely contributed to the black lid being separated from the housing.The root cause of this damage could not be determined.Penumbra engines are inspected at incoming quality control which includes a visual inspection as well as a verification of test results to ensure specifications for each output are met.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.Section h.Box 6.Conclusions code 1: 4316 - the investigation findings do not lead to a clear conclusion about the root cause of the damage noted on the engine.
|