Model Number M00510890
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, an alliance ii handle was used in conjunction with the trapezoid basket in an attempt to crush a 1.5 cm stone.However, the basket failed to crush the stone and the tip failed to separate from the basket to release the stone.The thumb ring was bent while attempting to crush the stone.The entrapped stone and basket was stuck in the duct and removed from the patient using an olympus emergency lithotripter device.The procedure was completed successfully with a different device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key11624371
MDR Text Key244726556
Report Number3005099803-2021-01481
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/27/2021
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0026253519
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2021
Initial Date FDA Received04/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1