Model Number M00510890 |
Device Problems
Break (1069); Separation Failure (2547)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a trapezoid rx basket was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, an alliance ii handle was used in conjunction with the trapezoid basket in an attempt to crush a 1.5 cm stone.However, the basket failed to crush the stone and the tip failed to separate from the basket to release the stone.The thumb ring was bent while attempting to crush the stone.The entrapped stone and basket was stuck in the duct and removed from the patient using an olympus emergency lithotripter device.The procedure was completed successfully with a different device.There were no patient complications reported as a result of this event.
|
|