The initial reporter stated they had issues with recovery for many controls run on the cobas 6000 e 601 module on (b)(6) 2021.It was determined the test counts for the measuring cells were high.The field service engineer adjusted the photomultiplier tube high voltage, ran system volume checks, and ran a blank cell calibration.On (b)(6) 2021, the customer stated they received discrepant results for two patient samples tested with elecsys ferritin on the e 601 analyzer.The reagent pack that was in use was replaced with a standby reagent pack and the samples were repeated, yielding different results.No incorrect results were reported outside of the laboratory.The first sample initially resulted in a ferritin value of 809.1 ng/ml, which repeated as 428.19 ng/ml.The second sample initially resulted in a ferritin value of 87.63 ng/ml, which repeated as 46.53 ng/ml.The serial number of the e 601 analyzer is (b)(4).
|
The reagent pack in use and the standby pack were both calibrated on (b)(6) 2021.The pack that was first in use had calibration signals noticeably lower than expected.The calibration signals from the standby pack were within expectations.Controls run on (b)(6) 2021 were within specified ranges.Upon review of the alarm trace, several abnormal aspiration alarms occurred on the day of the event.This could indicate issues related to sample quality, but likely not related to the event.The investigation determined the issue was consistent with poor calibration on the reagent pack that was first in use.A general instrument issue is unlikely.
|