• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN STEM; PROSTHESIS HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN STEM; PROSTHESIS HIP Back to Search Results
Catalog Number N
Device Problems Corroded (1131); Material Erosion (1214); Unstable (1667); Naturally Worn (2988)
Patient Problems Ossification (1428); Erosion (1750); Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Osteolysis (2377); Osteopenia/ Osteoporosis (2651); Joint Laxity (4526); Metal Related Pathology (4530); Insufficient Information (4580)
Event Date 07/06/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant medical products: part # unknown / unknown head/ lot # unknown.Part # unknown / unknown shell/ lot # unknown.Multiple mdr reports were filed for this event, please see associated reports: (b)(4).
 
Event Description
It was reported that patient underwent initial left total hip arthroplasty approximately 16 years ago.Subsequently, the patient has reported mild unknown symptoms and elevated chromium levels.However, no revision procedure has been reported to date.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
Reported event was unable to be confirmed.Device history record (dhr) review was unable to be performed as the item and lot number of the device involved in the event is unknown.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: summarized by hcp findings: recent metal ion levels cobalt 2.1 (normal <4.0), chromium 2.1 (normal <0.6) the left hip only has minor symptoms at this time root cause could not be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that patient underwent a left hip revision surgery 15 years post implantation due to metal related pathology.Pain and instability was reported.Surgeon noted surgery was extensive due to muscle damage, osteolysis, bone loss/erosion, ho and scar tissue attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues recent ion levels were high but lowered after the revision, extensive scar tissue, bone loss, muscle detachment, instability, wear debris, tumor, osteolysis, and altr as reported.Pain increase with weight-bearing.Corrosion on the head and taper, ho around the greater trochanter.No further complications were noted.The additional information does not change the outcome of the previous investigation.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN STEM
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11636309
MDR Text Key261203509
Report Number0001825034-2021-01013
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2021
Initial Date FDA Received04/08/2021
Supplement Dates Manufacturer Received04/09/2021
10/14/2021
11/10/2021
Supplement Dates FDA Received04/13/2021
11/08/2021
11/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Other;
Patient SexMale
Patient Weight104 KG
-
-