Model Number A219
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Failure to Sense (1559); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 01/09/2021
Event Type  Injury  
Event Description
It was reported that the patient received two inappropriate shocks due to oversensing of noise with decreased sensing.The field representative noted that the sensing appeared flat.The patient has no sternal wires and good measurements.Boston scientific technical services (ts) was contacted and discussed possible causes including air entrapment.Ts further suggested an x-ray.The physician agreed that this could be related to air entrapment and had the subcutaneous implantable cardioverter defibrillator (s-icd) programmed off for three weeks.The patient was brought in three weeks later where the set up was re-initiated and no noise was observed.Sensing was also good.The s-icd was turned back on and no additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11638758
MDR Text Key244615278
Report Number2124215-2021-04716
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/12/2021
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number248832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2021
Initial Date FDA Received04/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/24/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-