LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred during maintenance in (b)(6).The failure was traced back to a defective processor board.The part was replaced and the problem was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message associated to discrepancies between the temperature measured by the safety and control sensors during maintenance.There was no patient involvement.
|
|