MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71364073
Device Problems Break (1069); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2021
Event Type  malfunction  
Event Description
It was reported that, the internal piece of a mi z handle acet reamer was returned detached/broken.It is unknown whether the event happened during surgery and if there was a patient involvement and no further information is available.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the device is broken in four pieces.All pieces were returned.The device shows signs of extreme wear and tear.The manufacture date is 2009.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MI Z HANDLE ACET REAMER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11642780
MDR Text Key244691943
Report Number1020279-2021-02810
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010507570
UDI-Public03596010507570
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71364073
Device Catalogue Number71364073
Device Lot Number09CM03682
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/15/2021
Initial Date FDA Received04/09/2021
Supplement Dates Manufacturer Received04/19/2021
Supplement Dates FDA Received05/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-