Model Number WCD 4000
Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Date 02/28/2021
Event Type  Injury  
Manufacturer Narrative
Electrode belt sn (b)(4) was returned and evaluated at the distributor, in accordance with procedures recommended by zoll manufacturing corporation.The evaluation included review of downloaded software flag files (attached) on the day of the event and incoming functional testing.The review of the software flags consisted of an analysis of the downloaded data to identify any fault flags or unusual patterns of software flags.The software flag files did not suggest a device malfunction that would contribute to the inappropriate treatment.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.The pulse delivery circuitry test verified proper delivery of a full energy 150j biphasic pulse.The functional testing confirmed proper ecg acquisition and pulse delivery functionality.Monitor sn (b)(4) was returned to the distributor for evaluation.The evaluation of the monitor confirmed the service code 104/dl code 203.The sd card was defective.The distributor evaluation included downloaded data review as well as incoming functional testing of the device's ecg acquisition and pulse delivery circuitry, as recommended by zoll manufacturing.The monitor passed essential performance testing.The root cause of the defective sd card could not be positively identified.
 
Event Description
A us distributor contacted zoll to report that a patient experienced a defibrillation event consisting of one shock on (b)(6) 2021.The patient's ecg rhythm at the time of the treatment is unknown.The response buttons were pressed after the treatment was delivered.The patient continued wearing the lifevest.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key11645949
MDR Text Key245925159
Report Number3008642652-2021-02998
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/11/2021
Initial Date FDA Received04/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
-
-