• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCORD DUAL ENDED HEX DRIVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCORD DUAL ENDED HEX DRIVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71360046
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2021
Event Type  malfunction  
Event Description
It was reported that, during set up or inspection, the accord dual ended hex driver is stripped from prior use.There was no delay and procedure was completed with the same device.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The keyed tip of the hex screwdriver is worn with dull edges rendering the device inoperable.The device was manufactured in 2013 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed item.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD DUAL ENDED HEX DRIVER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646931
MDR Text Key244857005
Report Number1020279-2021-02890
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010611543
UDI-Public03596010611543
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360046
Device Catalogue Number71360046
Device Lot Number13AM06335
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/17/2021
Initial Date FDA Received04/11/2021
Supplement Dates Manufacturer Received04/09/2021
Supplement Dates FDA Received04/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-