MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA SINGLE USE GUIDWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA SINGLE USE GUIDWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number OL-XA25455
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2021
Event Type  malfunction  
Manufacturer Narrative
The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The production lot number was not provided by the user facility, which prevented a meaningful review of the device history record, and product-release judgement records, and product shipping inspection record.Regarding the involved product code, a search of the complaint file for the last two years found no other similar report from other facilities.Ifu states: if any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the guide wire and/or endotherapy accessory and determine the cause by fluoroscopy or endoscope.Continuing to manipulate the guidewire could cause patient injury, such as punctures, hemorrhages or mucous membrane damage.It may also damage the endoscope, instrument and/or endotherapy accessory.It is likely that when passing the severe stenosis, might have been exposed to an external load such as a pushing or bending load for some reason and damaged, which resulted in sticking inside the enbd tube.However, with no device return the exact cause of the reported event cannot be definitively determined based on the available information.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
 
Event Description
The user facility reported that the single use guidewire was used during the placement of enbd tube procedure.An attempt to insert a 7fr tube over a guidewire was made although failed due to severe stenosis.It was exchange for 5fr and insertion was done, at that time, the guidewire became impossible to be pulled out.They found that the guidewire was kinked when removed it together with the tube unwillingly.The procedure was completed successfully.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE GUIDWIRE
Type of Device
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
950 elkton blvd
elkton, MD 21921
8002837866
MDR Report Key11663049
MDR Text Key245328991
Report Number9681834-2021-00061
Device Sequence Number1
Product Code OCY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K091417
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberOL-XA25455
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/01/2021
Initial Date FDA Received04/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
JF-260V
-
-