• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA2QQ
Device Problem Misconnection (1399)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, the cardiac resynchronization therapy defibrillator (crt-d) exhibited an initial issue inserting the right ventricular (rv) lead defibrillation lead into the port of the header.The lead would not fully insert to show the blue line that indicated full insertion.The pin had some bunching due to grip and attempting to insert the plastic grip prior to the pin portion of the lead.The pin was repeatedly cleaned and the set screw within the header was slightly loosened.After recleaning the lead pin with saline, the lead was able to be fully inserted and showed the blue line indicator.The device and rv lead remain in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11664408
MDR Text Key245353797
Report Number9614453-2021-01550
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2022
Device Model NumberDTMA2QQ
Device Catalogue NumberDTMA2QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2021
Initial Date FDA Received04/14/2021
Date Device Manufactured11/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, 6935M62 LEAD, 479888 LEAD
Patient Age73 YR
Patient Weight75
-
-