Model Number 8360-10
Device Problems Material Separation (1562); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap inc.That a prestige atra grasper dbl-act 5mm (part # 8360-10) was in need of repair.According to the complainant, the distal tip separated from the shaft.The failure was observed in sterile processing.The complaint device was returned to the manufacturer for evaluation.No patient involvement.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Manufacturing site evaluation: aesculap inc.Previously reported that a supplier corrective action request (scar) was initiated due to an adverse trend observed for devices exhibiting failure at the thumb-loop assembly joint.The supplier evaluated the malfunction which revealed an issue with the push rod fixture.As a result of their findings, the push rod fixture was redesigned to increase the clearance, which ensured that the fixture appropriately stressed the entirety of the soldering joint.Upon implementing the new fixture, the supplier tested returned non-conforming samples, and verified that the soldering no longer hung up on the new fixture, as had been observed on the previous one the complaint device was returned to the manufacturer for physical evaluation.A visual examination was performed which confirmed a separation of the distal tip.No visible defects to the proximal weld were observed.An investigation of the device manufacturing records was conducted by the manufacturer for the lot # of the device in question.No non-conformances were reported.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Additionally, historical scrap rates were reviewed with no increase observed in scrap related to the complaint issue.In addition to the redesign of the soldering fixture, the supplier reviewed the work instruction for the torch soldering operation and identified improvement opportunities.The original work instruction was used to better define the soldering process with a more focused emphasis on the following: equipment startup and shutoff operations, clear imagery of acceptable soldered subassemblies, and clarified cleaning operations for components prior to soldering.A second dedicated work instruction was implemented to better define the attribute inspection criteria for the brazing process used for the solder between the thumb loop and rotator block.The investigation into the cause of the reported problem was able to confirm the failure mode of a distal tip separation.This event likely occurred due to inadequacies in the defined production process which limited the device performance.Therefore, the most probable root cause is considered to be manufacturing related.In addition to a supplier corrective action request (scar) being initiated, a corrective action/preventive action (capa) was opened by aesculap inc.For further evaluation of the design transfer of this device.
 
Event Description
No update needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
MDR Report Key11667884
MDR Text Key262726309
Report Number2916714-2021-00068
Device Sequence Number1
Product Code NWV
UDI-Device Identifier04046955083374
UDI-Public4046955083374
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8360-10
Device Catalogue Number8360-10
Device Lot NumberL50429911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2021
Initial Date Manufacturer Received 03/15/2021
Initial Date FDA Received04/14/2021
Supplement Dates Manufacturer Received10/20/2021
Supplement Dates FDA Received10/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-