• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT M.R.I. IMPLANTABLE 8F GROSHONG SINGLE-LUMEN; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT M.R.I. IMPLANTABLE 8F GROSHONG SINGLE-LUMEN; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1878000
Device Problems Fracture (1260); Material Separation (1562); Migration (4003)
Patient Problems Pulmonary Embolism (1498); Cardiac Perforation (2513)
Event Date 05/30/2019
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 01/2021).
 
Event Description
It was reported that post port device implant, the catheter was allegedly fractured and the patient underwent surgery to remove the fractured catheter from the right ventricle.It was further reported that the device embolized the right atrium.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: this is the first complaint reported to date for this product and lot, therefore a device history record review is not required.Investigation summary: the sample was not returned for evaluation, a medical record review was performed.The investigation is confirmed for the reported catheter fracture, separation, and embolism, as the medical record review identified catheter tubing lodged within the coronary sinus extending into the right atrium.According to the medical record review, all known catheter fragments were retrieved and the retrieval of the fragments was uneventful.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Instructions for use reviewed.The instructions for use states that, "this device is contraindicated for catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates of pinch-off." ¿warning: do not power inject through a port system that exhibits signs of clavicle-first rib compression or pinch-off, as it may result in port system failure.¿ ¿signs of pinch-off clinical: difficulty with blood withdrawal resistance to infusion of fluids patient position changes required for infusion of fluids or blood withdrawal radiologic: grade 1 or 2 distortion on chest x-ray.Pinch-off should be evaluated for degree of severity prior to explantation.Patients indicating any degree of catheter distortion at the clavicle/first rib area should be followed diligently.There are grades of pinch-off that should be recognized with appropriate chest x-ray as follows¿ ¿preventing pinch-off the risk of pinch-off syndrome can be avoided by inserting the catheter via the internal jugular vein (ij).Subclavian insertion of the catheter medial to the border of the first rib may cause catheter pinch-off, which in turn results in occlusion causing port system failure during power injection.If you choose to insert the catheter into the subclavian vein, it should be inserted lateral to the border of the first rib or at the junction with the axillary vein because such insertion will avoid compression of the catheter, which can cause damage and even sever the catheter.The use of image guidance upon insertion is strongly recommended.A radiographic confirmation of catheter insertion should be made to ensure that the catheter is not being pinched.¿ h10: b5, d4 (expiry date: 01/2021).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported through the litigation process that post port device implant, the catheter was allegedly fractured and the patient underwent surgery to remove the fractured catheter from the right ventricle.It was further reported that the device embolized the right atrium.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT M.R.I. IMPLANTABLE 8F GROSHONG SINGLE-LUMEN
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11672191
MDR Text Key245599987
Report Number3006260740-2021-01340
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027246
UDI-Public(01)00801741027246
Combination Product (y/n)N
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1878000
Device Catalogue Number1878000
Device Lot NumberREBX2043
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/19/2021
Initial Date FDA Received04/15/2021
Supplement Dates Manufacturer Received07/01/2021
Supplement Dates FDA Received07/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age70 YR
-
-