Review of the device history records did not identify any manufacturing or processing related irregularities.The invictus mis reduction polyaxial screwdriver, locking was found to be properly manufactured and released in accordance with design specifications.The mis reduction polyaxial screwdriver, locking returned to atec on 03/24/2021.Visual inspection indicates a clean shear fracture of the driver hexalobe tip.No other remarkable features or observations were identified.The root cause is likely physical damage due to excessive torque which resulted in the tip shearing off.No patient injury reported.Labeling review: "possible adverse effects: the following complications and adverse reactions have been shown to occur with the use of similar spinal instrumentation.These effects and any other known by the surgeon must be discussed with the patient preoperatively.1.Initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.".
|