• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SPIN REDUCTION FORCEPS - BOWED 205MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SPIN REDUCTION FORCEPS - BOWED 205MM Back to Search Results
Model Number 71175088
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Event Description
It was reported that two spin reduction forceps - bowed 205 mm were noticed with the tips point bent.No case involved therefore no patient impact.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIN REDUCTION FORCEPS - BOWED 205MM
Type of Device
FORCEPS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11683501
MDR Text Key245998931
Report Number1020279-2021-03074
Device Sequence Number1
Product Code HTD
UDI-Device Identifier00885556636718
UDI-Public00885556636718
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71175088
Device Catalogue Number71175088
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/23/2021
Initial Date FDA Received04/18/2021
Supplement Dates Manufacturer Received06/04/2021
Supplement Dates FDA Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-