Model Number 74011711
Device Problems Difficult to Remove (1528); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka procedure, it was different to thread up and down with the journey ii cr lock fem implant impactor; physician had a hard time getting the implant off while inside the patient.There was a slight delay of less or equal to 30 minutes.Procedure was completed with a backup device.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR LOCK FEM IMPLANT IMPACTOR
Type of Device
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11683528
MDR Text Key246000606
Report Number1020279-2021-03101
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556422892
UDI-Public00885556422892
Combination Product (y/n)N
PMA/PMN Number
K121443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74011711
Device Catalogue Number74011711
Device Lot Number16CBG0034
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/23/2021
Initial Date FDA Received04/18/2021
Supplement Dates Manufacturer Received07/08/2021
Supplement Dates FDA Received07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-