OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information is not yet available for this event.Event date is not known.Supplemental report(s) will be filed as any information becomes available.The device has been returned and a device evaluation completed for it.Manufacture date is not known.Upon inspection and testing, the device passed inspection.The user¿s complaint could not be duplicated.The device had some missing screws from the rear panel and one from the heat sink.Also, the device had a third-party lamp.
 
Event Description
As reported for this event, during preparation for use, the e102 lamp error was observed for the device.There is no patient involvement and no harm reported for any patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The following sections were updated: g3, g6, h2, h4, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon (e102 error) was not reproduced during inspection, however, this may have occurred due to the use of a non-olympus lamp.A review of the instruction manual identifies the following verbiage regarding the replacement of the lamp: "6.1 replacement of the examination (xenon) lamp - always use the examination lamp designated below.To order a new examination lamp, contact olympus.Xenon lamp md-631.Never install a lamp that has not been approved by olympus.The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11684465
MDR Text Key268928228
Report Number8010047-2021-05101
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/25/2021
Initial Date FDA Received04/19/2021
Supplement Dates Manufacturer Received05/20/2021
Supplement Dates FDA Received05/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-