• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Model Number IPN913776
Device Problem Failure to Sense (1559)
Patient Problem Insufficient Information (4580)
Event Date 03/29/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of the reported failure mode was conducted and 8 devices were taken from current production (425-00 concha neptune lot # 73c210010n) at the manufacturing facility and were functionally tested.No issues were encountered during the functional testing.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported "the heater humidity gradient was not working".Unknown where alleged issue was found at time of report.Unknown patient condition at time of report.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no major defects observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was able to power on and off with no issues.The unit was then prepared for the functional bench test where water, an adult breathing circuit (880-36kit), 2-3 lpm of air pressure, dual temperature probe, sand a 382-10 concha smart water column was connected to the unit for a real time operational scenario.The unit successfully negotiated all pre-operational self-tests again and was functioning in real time.All settings were able to be adjusted without any issues.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
It was reported "the heater humidity gradient was not working".Unknown where alleged issue was found at time of report.Unknown patient condition at time of report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key11685279
MDR Text Key246012263
Report Number3003898360-2021-00358
Device Sequence Number1
Product Code BTT
UDI-Device Identifier14026704646811
UDI-Public14026704646811
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberIPN913776
Device Catalogue Number425-00
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Initial Date Manufacturer Received 03/29/2021
Initial Date FDA Received04/19/2021
Supplement Dates Manufacturer Received05/18/2021
Supplement Dates FDA Received05/18/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-