Manufacturer Narrative
The biomedical engineer reported that they are having intermittent signal loss issues and artifacts/noise occurring only in one specific department.The biomedical engineer states it is happening in the four (4) east wing and is isolated to just that location.Nihon kohden service engineer went on-site and tightened up three (3) way splitters and aac connections.All antenna's are working as expected.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.(b)(4).
 
Event Description
The biomedical engineer reported that they are having intermittent signal loss issues and artifacts/noise occurring only in one specific dept.The biomedical engineer states that it is happening in the four (4) east wing and is isolated to just that location.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9110A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
MDR Report Key11694237
MDR Text Key247127385
Report Number2080783-2021-00237
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103906
UDI-Public04931921103906
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9110A
Device Catalogue NumberORG-9110A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/20/2021
Distributor Facility Aware Date03/23/2021
Event Location Hospital
Date Report to Manufacturer04/20/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-