MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. HCG ONE STEP PREGNANCY TEST STRIP (URINE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. HCG ONE STEP PREGNANCY TEST STRIP (URINE) Back to Search Results
Model Number FHC-101
Device Problem False Negative Result (1225)
Patient Problem Ectopic Pregnancy (1819)
Event Date 02/09/2021
Event Type  Injury  
Manufacturer Narrative
Customer unwilling to return.Investigation conclusion: the case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.Product testing and manufacturing record review could not be performed as a lot number could not be obtained and product could not be returned.A root cause could not be determined from the available information.However, case details indicate that the result was read at 5-10 minutes, and a timer was not used.Per the directions for use, read the result at 3-4 minutes.Do not interpret results after the appropriate read time.A timer is required.Complaints are tracked and trended on a monthly basis.Per the package insert, this test may produce false negative results.False negative results may occur when the levels of hcg are below the sensitivity level of the test.When pregnancy is still suspected, a first morning urine specimen should be collected 48 hours later and tested.In case pregnancy is suspected and the test continues to produce negative results, see a physician for further diagnosis.
 
Event Description
(b)(6) 2021: patient presented to facility after obtaining a positive result on a home pregnancy test (exact brand/specificity not provided).Urine specimen was tested on the hcg one step pregnancy test strip (urine) kit and a negative result was obtained.Due to the negative result, the patient was incorrectly discharged from the hospital.(b)(6) 2021: the patient suffered a ruptured fallopian tube at home and urgent surgical intervention was carried out.Patient had an ectopic pregnancy rupture with intra-abdominal bleeding.The procedure performed was urgent abdominal surgery with unilateral salpingectomy.Confirmatory beta hcg test provided a positive result of 7200 miu/ml.Customer requested they not be contacted further regarding this event.No further information able to be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HCG ONE STEP PREGNANCY TEST STRIP (URINE)
Type of Device
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
monika burrell
9975 summers ridge road
san diego, CA 92121
8588052506
MDR Report Key11700565
MDR Text Key262866797
Report Number2027969-2021-00033
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K993203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-101
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/13/2021
Initial Date FDA Received04/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age25 YR
-
-