• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 74122546
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Metal Related Pathology (4530)
Event Date 08/14/2019
Event Type  Injury  
Event Description
Us legal.It was reported that the patient underwent a medically indicated revision of the bhr-tha hybrid system right hip on (b)(6) 2019 due to pain, limited mobility and elevated ion levels.Among the intraoperative findings there was dark brown, dark gray fluid, evidence of metallosis and corrosion between the femoral head and the trunnion.The patient outcome is unknown.
 
Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.As no device batch numbers were provided for investigation, a manufacturing record, complaint history and device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.Based on the reported symptoms it cannot be concluded that the events/clinical reactions (pain, limited mobility, elevated ion levels and metallosis) were associated with a mal-performance of the implant.The patient impact cannot be determined at this time.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, our investigation remains inconclusive and a definitive root cause cannot be determined.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 46MM
Type of Device
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key11708253
MDR Text Key246745843
Report Number3005975929-2021-00190
Device Sequence Number1
Product Code KWL
UDI-Device Identifier00885556071588
UDI-Public885556071588
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number74122546
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/30/2021
Initial Date FDA Received04/22/2021
Supplement Dates Manufacturer Received05/12/2021
Supplement Dates FDA Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-