MEDTRONIC, INC. ACHIEVE ADVANCE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 2ACH20 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the 2ach20 mapping catheter with lot 5965875 was returned and analyzed.Visual inspection showed that the pebax tubing was detached from the shaft and was ribbed at multiple locations along the pebax, starting at the transition with the metal shaft to the loop array.In conclusion, the mapping catheter failed the returned product inspection due to a broken pebax tube, displaced electrode, and ribbed pebax.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
After a completed case, the mapping catheter subsequently tested out of specification per the manufacturers investigation.
|
|