A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the drive manifold, and performed a full preventive maintenance with all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
Updated fields: b4, b6, b7, d4(model#), d11, g4, g7, h2, h6(investigation type), h10, h11 corrected fields: d1, d5, g1(contact person) h4, h6(problem code) analysis of production: (3331/102) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/102) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/102) the overall 24 month product complaint trend data for the period apr 2019 through mar 2021 was reviewed.There were no triggers identified for the review period.
|