• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problems Electrical /Electronic Property Problem (1198); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the drive manifold, and performed a full preventive maintenance with all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) had an electrical error.It is unknown the circumstances under which the event occurred.However, there was no patient involvement, and no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, b6, b7, d4(model#), d11, g4, g7, h2, h6(investigation type), h10, h11 corrected fields: d1, d5, g1(contact person) h4, h6(problem code) analysis of production: (3331/102) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/102) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/102) the overall 24 month product complaint trend data for the period apr 2019 through mar 2021 was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11710969
MDR Text Key247292436
Report Number2249723-2021-00850
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2021
Initial Date FDA Received04/22/2021
Supplement Dates Manufacturer Received03/07/2022
Supplement Dates FDA Received03/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2002
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-