|
Model Number H701374 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
One viewmate¿ z catheter interface module was received for analysis.Visual inspection of the returned module indicated signs of normal wear consistent with clinical use.No physical damage was detected to either connector end or module electrical terminals.Functional testing revealed the ultrasound console did not detect the catheter, which confirmed the field reported issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Based on the information provided to abbott and the investigation performed, the root cause of the field reported issue was isolated abnormal functionality the catheter detection circuitry.
|
|
Event Description
|
During a cryo atrial fibrillation ablation procedure, the cim did not recognize the catheters.A transesophageal echocardiogram (tee) was performed instead with anesthesia administered to complete the procedure successfully with no consequences to the patient.
|
|
Search Alerts/Recalls
|
|
|