BAXTER HEALTHCARE CORPORATION TITANIUM TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number T5C4326K |
Device Problems
Loose or Intermittent Connection (1371); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the connection between the patient adaptor of a transfer set assembly was loose and could not connect to the patient¿s titanium adaptor.This occurred during use of the device at the hospital for peritoneal dialysis therapy.There was no allegation against the titanium adaptor and it was still in use.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|