MEDTRONIC NEUROMODULATION VERIFY ENHANCED; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 353101 |
Device Problems
Display or Visual Feedback Problem (1184); Migration or Expulsion of Device (1395)
|
Patient Problems
Dizziness (2194); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: neu_unknown_lead, lot#: unknown, implanted: (b)(6) 2021, product type: lead.Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from an advanced evaluation trial patient who was using an external neurostimulator (ens) for non-obstructive urinary retention.The trial began (b)(6) 2021.It was reported that the patient's wires came loose.They were worried the therapy was not working as there was no light on the device on their back.The issue was resolved at the time of the report.
|
|
Manufacturer Narrative
|
Continuation of d10: product id neu_unknown_lead lot# unknown serial# implanted: (b)(6) 2021 explanted: product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the patient.They reported that they were light-headed after the last therapy adjustment, on the previous call.
|
|
Search Alerts/Recalls
|
|
|