Other, other text: h6: event problem and evaluation codes: updated after device evaluation h10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection: no occlusions nor obstructions were detected on the samples.Functional test: the complaints sample was tested using the cadd pump solis [cal.Id: 1.0326; due date: august, 2021] to prime the devices as instruction of use as10003589-004 rev.100 indicate.No difficulty were detected during the priming and no alarms were activated, the water could pass through the whole devices; thus the failure mode reported is not confirmed.The cause of the reported problem could not be determined.No corrective actions are required since the complaint was not confirmed.
|