• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 SROM 9/10 16X11X130 30; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 SROM 9/10 16X11X130 30; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 900531210
Device Problem Difficult to Remove (1528)
Patient Problem Insufficient Information (4580)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
On (b)(6) 2021, the revision surgery was performed.In the revision, after the head was removed, the surgeon tried to remove the stem, but the stem couldn¿t be removed, and it was left in the patient.The competitor implants were set, and the surgery was completed with a 30 minute delay.The surgeon commented that the cold welding occurred between the stem and sleeve and the working space was narrow.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM 9/10 16X11X130 30
Type of Device
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key11732089
MDR Text Key247509064
Report Number1818910-2021-08869
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number900531210
Device Lot Number7797527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2021
Initial Date FDA Received04/27/2021
Supplement Dates Manufacturer Received06/07/2021
05/13/2022
Supplement Dates FDA Received06/09/2021
05/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
S-ROM*SLEEVE PRX ZTT, 16B-LRG; SROM 9/10 16X11X130 30; S-ROM*SLEEVE PRX ZTT, 16B-LRG; SROM 9/10 16X11X130 30
Patient Age71 YR
Patient SexFemale
Patient Weight68 KG
-
-